PROVA-8000 Programmable Power Supply |
|
Over-Voltage , Over-Current Protection Bulit-in Buzzer Alarm Easy RECALL and STORE functions Stby / Operate Button English-like RS-232C commands OCP With Programmable Timeter 200 Sets of Memory AUTO-STEP function with proggrammable interval Bi-direction RS-232C interface 3.8 kg / 8.4 Lbs weight
|
|
|
|
Programmed Voltage and Current (5 minutes after power on) |
|
Range |
Resolution |
Accuracy |
Comments |
Voltage V (C.V.) |
0 – 40V |
1mV
|
0.05% ± 9mV |
Typical ± 5mV |
Current A (C.C.) |
0 –5V |
1mA |
0.2% ± 9mA |
Typical ± 5mA |
C.V.: Constant Voltage C.C.: Constant Current | |
Voltage and Current Readback |
|
Range |
Resolution |
Accuracy |
Voltage V (C.C.) |
0 – 40V |
1mV
|
0.1% ± 12mV |
Current A (C.C.) |
0.05 –5A |
1mA |
0.2% ± 12mA |
C.V.: Constant Voltage C.C.: Constant Current | |
Load Regulation |
1mV/A |
Line Regulation |
No effect if line voltage falls within 90V to 130V for 115V nominal, 200V to 240V for 230V nominal. |
Fixed 5V / 2A Accuracy |
5V ± 0.25V |
Fixed 3.3V / 3A Accuracy |
3.3V ± 0.16V |
: When 5V or 3.3V output is less than 1V (shorted), the corresponding symbol will be shown in LCD to warn the users. | |
Ripple and Noise (0 ~ 40V, 0 ~ 5A) |
<3mVrms (C.V.), typical 1mVrms |
Ripple and Noise (0 ~ 40V, 0 ~ 5A) |
<3mArms (C.C.), typical 1mArms |
Ripple and Noise (5V/2A, 3.3V/3A) |
10mVrms ~ 20mVrms, 10mArms ~20mArms |
Temperature Coefficient |
< 100ppm/℃ |
OCP Detection |
Current A > 0.05A |
C.V. or C.C. Detect and Display |
Current > 0.05A |
Power Source |
115V (90V to 130V) or 230V (200V to 240V) |
Accessories |
power cord x 1, users manual x 1 |
Dimensions |
310mm x 250mm x 133 mm / 12.2" x 9.8" x 5.2" |
Weight |
3.8kg / 8.4lbs | | |
|
LANGUAGUE |
|
| |