PROVA-133 Process Calibrator/Simulator + Function Generator
Simulate 4~20 mA process transmitter Source 4~20 mA loop Current Simulate electronic load (Max. 30V, 20mA) Test LED brightness (0~24mA) Source -9.999~70mV and 0~15V Calibrate temperature of 11 types of thermocouples Handheld function generator of sine wave, square wave, triangular wave, arbitrary wave from 0.3 to 20KHz)
Product Features
Simulate 4~20 mA process transmitter
Source 4~20 mA loop Current
Simulate electronic load (Max. 30V, 20mA)
Test LED brightness (0~24mA)
Source -9.999~70mV and 0~15V
Calibrate temperature of 11 types of thermocouples
Handheld function generator of sine wave, square wave, triangular wave, arbitrary wave from 0.3 to 20KHz)
Pulse generator (3μS to 999.99mS)
Map 4~20 mA into engineering units
Source: mA, V, Hz, sine wave, square wave, triangular wave, truncated sine wave, user programmable waveform and temperature.
Measure: Current (DC mA), Voltage (DC mV, DC V), and temperature (°C, °F)
Measure 4~20 mA with 24V loop supply simultaneously
Selectable 250Ω HART resistor to facilitate use with HART communication device
Programmable cold junction compensation for temperature measurement
Auto step and auto ramp for sourcing mA, V, and temperature
Dot matrix LCD with backlight
Rechargeable lithium battery (1600mAh)
Data logging function for source and measurement
Program calibrator through PC USB port
Programmable 0% and 100% value for easy 25% step function
DTMF (Dual Tone Multi-Frequency) can perform professional testing for telephone line and audio product (MP3 or MD)
Electrical Specifications: (23+/- 5℃, 10 minutes after turning on the power)
mA (source) (Vopen > 24V)
Range
Resolution
Accuracy of Reading
0.005mA to 4mA
1uA
+/-0.03% +/-5dgts
4mA to 20mA
+/-0.03% +/-3dgts
20mA to 24mA
+/-0.03% +/-5dgts
V (source) (maximum load 1mA, short circuit protection < 100mA)
Range
Resolution
Accuracy of Reading
0.005V to 10V
0.001V
+/-0.03% +/-5dgts
10V to 15V
mA (measure)
Range
Resolution
Accuracy of Reading
-4mA to -0.005mA
1μA
+/-0.03% +/-10dgts
0.005mA to 4mA
+/-0.03% +/-5dgts
4mA to 20mA
+/-0.03% +/-3dgts
20mA to 24mA
+/-0.03% +/-5dgts
If reading of mA (measure) is less than 5 digits, it is displayed as 0.
V (measure)
Range
Resolution
Accuracy of Reading
-3V to -0.005V
0.001V
+/-0.03% +/-10dgts
0.005V to 10V
+/-0.03% +/-5dgts
10V to 24V
+/-0.03% +/-5dgts
If reading of V (measure) is less than 5 digits, it is displayed as 0.